The international calibration procedure for B3 film dosimetry system to ensure the quality irradiated products by 10 MeV electron beam accelerators at VINAGAMMA

Pham Thi Thu Hong1, Nguyen Thi Ly1, Nguyen Thanh Duoc1, Doan Binh1, Cao Van Chung1, Doan Thi The1
1 Research and Development Center for Radiation Technology, 202A street 11, Linh Xuan ward, Thu Duc district, HCM. City, Vietnam

Main Article Content

Abstract

We performed an in-plant calibration of dosimetry system by electron beam (EB) irradiating the B3 film dosimeters at VINAGAMMA, and inter-compared with the alanine dosimetry, which were supplied and analyzed by Risø High Dose Reference Laboratory (HDRL) as the reference standard. The results revealed that the relative deviation between the values of absorbed doses obtained with our dosimeter and the transfer standards dosimeter measured by HDRL was within the acceptable limitation (about ± 3.0 % in the target range of 2.0-10.0 kGy). And post-irradiation stability of B3 film dosimeters was still maintained after 180 days storage. It is suggested that the B3 film dosimetry could be used in routine radiation processing at VINAGAMMA with the investigated dose range for quality assurance of the irradiated products, specially are foods and foodstuffs processed under the 10 MeV EB accelerator at VINAGAMMA.

Article Details

References

[1] ASTM International (2nd edition), “Standard practice for calibration of routine dosimetry systems for radiation processing”, Standards on Dosimetry for Radiation Processing, 23-41, 2004.
[2] P. Sharpe and A. Miller, “NPL CIRM Report 29 - Guidelines for the calibration of dosimeters for use in radiation processing”, National Physical Laboratory, UK, 2009.
[3] GEX Corporation USA, “B3 radiochromic film dosimetry”, 2010.
[4] M. Bazioglou and J. Kalef-Ezra, “Dosimetry with radiochromic films: a document scanner technique, neutron response, applications”, Applied Radiation and Isotopes, 55, 339–345, 2001.
[5] C. G. Soares. “Radiochromic film dosimetry”, Radiation Measurements 41, S100–S116, 2007.
[6] GEX Corporation USA, “GEX DOC # 100 – 263 C - Performing a dosimeter batch calibration”, 2007.
[7] ASTM International (2nd edition), “Standard practice for use of a radiochromic film dosimetry system”, Standards on Dosimetry for Radiation Processing, 43-47, 2004.
[8] GEX Corporation USA, “GEX GEX DOC # 100-258 D - Guidelines for the measuring GEX dosimeters”, 2007.
[9] ASTM International (2nd edition), “Guide for estimation of measurement uncertainty in dosimetry for radiation processing”, Standards on Dosimetry for Radiation Processing, 173-193, 2004.