Establishment and quality assessment of X-ray narrow spectrum series according to ISO 4037:2019 at SSDL of Center For Nuclear Technologies

Le Huu Loi1, Nguyen Hoang Long1, Ong Quang Son1, Ho Manh Dung1, Dao Van Hoang1, Truong Truong Son2, Vo Van Tien1, Le Quang Vuong2
1 Center for Nuclear Technologies (CNT), 217 Nguyen Trai Street, District 1, Ho Chi Minh City, Vietnam
2 HCMC University of Education, 280 An Duong Vuong Street, District 5, Ho Chi Minh City, Vietnam

Main Article Content

Abstract

The X-ray narrow spectrum series in the range from 40 to 120 kV were established according to the requirements of the ISO 4037:2019. This study was conducted by experimental measurements and the use of the SpekPy v2.0 – a software toolkit for modeling of X-ray spectra. The experiments were done to determine inherent filtration (0.332 mm and 0.492 mm Al at 100 cm and 250 cm, respectively). The HVL values obtained by experiment and simulation were biased 0.10 and 0.09 mm Al, respectively as compared with ISO 4037:2019. The homogeneity coefficient h ranged from 0.91 to 0.97 by experiment and from 0.89 to 0.96 by simulation. Contribution of the scattering component was evaluated less than 4.0%. Air kerma rate values at distances of 100 cm and 250 cm have also been determined with the uncertainties ranging from 3.85% to 5.18% by experiment and 3.41% to 4.19% by simulation. The obtained results revealed that the X-ray radiation field established at CNT providing that it meets the requirements of the ISO 4037:2019, which is fitting for the purpose of calibration of survey meters and personal dosimeters.

Article Details

References

[1]. International Standard, ISO 4037-1:2019, Radiological protection - X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy - Part 1: Radiation characteristics and production methods, 2019.
[2]. International Standard, ISO 4037-2:2019, Radiological protection - X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy — Part 2: Dosimetry for radiation protection over the energy ranges from 8 keV to 1,3 MeV and 4 MeV to 9 MeV, 2019.
[3]. International Standard, ISO 4037-3:2019, Radiological protection - X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy - Part 3: Calibration of area and personal dosemeters and the measurement of their response as a function of energy and angle of incidence, 2019.
[4]. International Atomic Energy Agency, Calibration of radiation protection monitoring instruments, IAEA Safety Report Series No.16 (2000).
[5]. International Atomic Energy Agency, Measurement Uncertainty - A Practical Guide for Secondary Standards Dosimetry Laboratories, IAEA-TECDOC-1585, 2008.
[6]. G Poludniowski, A Omar, R Bujila and P Andreo, Technical Note: SpekPy v2.0 - a software toolkit for modeling X-ray tube spectra. Med Phys., 2021.
[7]. R Bujila, A Omar and G Poludniowski, A validation of SpekPy: a software toolkit for modelling x-ray tube spectra. Phys Med., 2020.
[8]. A. Herrati, M. Arib, T. Sidahmed, K. Khalal-Kouache, Establishment of ISO 4037-1 X-ray Narrow-Spectrum Series at SSDL of Algiers, Radiation Protection Dosimetry, pp. 1-18, 2016.
[9]. D. Zhang, W. Lai, J. Wu, H. Du, R. Zhao, S. Fan, Establishment of ISO 4037-1 X-ray narrow-spectrum series, J. Eng., Vol. 2019 Iss. 23, pp. 8858-8861, 2019.
[10]. Z. Qingli, H. Minchen, H. Yawen, W. Yingjing, N. Qiang, Establishment of Filtered X-ray Reference Radiation Based on ISO 4037-1:1996 in CIRP, Journal of Nuclear Science and Technology, Vol. 45:sup5, pp. 278-281, 2008.
[11]. L. Zhe, W. Jinjie, Z. Rui,S. Shengtao,W. Ji, L. Yunfei, Establishment of radiation in 10-40 kV narrow spectrum series, ICEMI'2019, 2019.
[12]. P. C. Franciscatto, M. D. P. A. Potiens, Determination of inherent and additional filtration in order to establish radiation qualities according to IEC 61267, INAC 2009, 2009.
[13]. Ong Quang Son, Le Huu Loi, Nguyen Hoang Long, Vo Van Tien, Determination of narrow spectrum qualities for X-ray standard system at secondary standard dosimetry laboratory, Center for Nuclear technologies, VIETNAM conference on nuclear science and technology (VINANST-14), 2021.
[14]. Le Huu Loi, Nguyen Hoang Long, Ong Quang Son, Le Quang Vuong, Truong Truong Son, Dao Van Hoang, The establishment and quality assessment of low-energy narrow-spectrum N-40 according to ISO 4037:2019, The 7th international conference on Applied and Engineering physics (CAEP7), 2021.