Self-characterization of irradiation facility using synthetic multi-element standard for determination of k0-factors of seven radionuclides of interest

Phonesavanh Lathdavong1,2, Ho Manh Dung3, Ho Van Doanh3, Tran Quang Thien4, Truong Truong Son5, Hoang Sy Minh Tuan6
1 Radiation and Nuclear Safety Office
2 Ministry of Education and Sports of Lao PDR
3 Center for Nuclear Technologies
4 Dalat Nuclear Research Institute
5 HCMC University of Education
6 Thu Dau Mot University

Main Article Content

Abstract

In this study, the k0 factors of seven short and medium-lived radionuclides (i.e. 66Cu, 52V, 38Cl, 134mCs, 128I, 140La, and 56Mn) were experimentally determined at the Dalat nuclear research reactor (DNRR). The synthetic multi-element standard (SMELS) samples were irradiated at channel 7-1 of the DNRR. SMELS Type III contains Au and Zr, thus it allows the self-characterization of the irradiation facility (i.e. the direct determination of f and a with only this material). The k0 factors of seven short and medium-lived radionuclides obtained in this work are in good agreement with the reference k0 factors. The ratio values between the experiment and reference k0 values were 66Cu of 1.004, 52V of 1.004, 38Cl of 0.961 and 0.976 (1642.7 keV and 2167.4 keV, respectively), 134mCs of 0.831, 128I of 1.022, 140La of 0.936, 0.955 and 0.955 (328.8 keV, 487.0 keV and 1596.2 keV, respectively), and 56Mn of 0.985 and 0.973 (846.8 keV and 1810.7 keV, respectively). The difference of k0 values between the experiment and the reference for the above radionuclides were lower than 7%. The present result indicates that the k0 factors can be determined by using the SMELS materials.

Article Details

Author Biographies

Ho Manh Dung, Center for Nuclear Technologies

217 Nguyen Trai Str., District 1, Ho Chi Minh City, 71014, Viet Nam

Ho Van Doanh, Center for Nuclear Technologies

217 Nguyen Trai Str., District 1, Ho Chi Minh City, 71014, Viet Nam

Tran Quang Thien, Dalat Nuclear Research Institute

01 Nguyen Tu Luc St., Dalat City, 67000, Viet Nam

Truong Truong Son, HCMC University of Education

280 An Duong Vuong St., Dist. 5, HCMC 72711, Viet Nam

Hoang Sy Minh Tuan, Thu Dau Mot University

Tran Van On Street, Phu Hoa Ward, Binh Duong, Vietnam

References

[1]. H.M. Dung, Development of k0-based cyclic neutron activation analysis for short-lived radionuclides, Journal of Radioanalytical and Nuclear Chemistry, 291, 485-492, 2012.
[2]. H.M. Dung, P.D. Hien, The application and development of k0-standardization method of neutron activation analysis at Dalat research reactor, Journal of Radioanalytical and Nuclear Chemistry, 257, 643-647, 2003.
[3]. e.a. Doanh HV, A new rapid neutron activation analysis system at Dalat nuclear research reactor, Nuclear Science and Technology, 4, 82-91, 2014.
[4]. H. Doanh, C. Vu, T. Thien, P. Son, N. Sy, N. Giang, N. Dien, A new rapid neutron activation analysis system at Dalat nuclear research reactor, Nuclear Science and Technology, 4, 84-91, 2014.
[5]. R. Acharya, A. Nair, A. Reddy, S. Manohar, Validation of a neutron activation analysis method using k0-standardization, Applied Radiation and Isotopes, 57, 391-398, 2002.
[6]. F. De Corte, A. Simonits, Recommended nuclear data for use in the k0-standardization of neutron activation analysis, Atomic Data and Nuclear Data Tables, 85, 47-67, 2003.
[7]. G.L. Molnar, Z. Revay, L. Szentmiklosi, New perspectives for very short-lived neutron activation analysis, Journal of Radioanalytical and Nuclear Chemistry, pp. 157-163, 2003.
[8]. M. Kubešová, J. Kučera, Validation of k0-standardization method in neutron activation analysis—the use of Kayzero for Windows programme at the nuclear physics institute, Řež, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 622, 403-406, 2010.
[9]. F. De Corte, A. Simonits, k 0-Measurements and related nuclear data compilation for (n, γ) reactor neutron activation analysis: IIIb: tabulation, Journal of Radioanalytical and Nuclear Chemistry, 133, 43-130, 1989.
[10]. R. Acharya, J. Holzbecher, A. Chatt, Determination of k0-factors of short-lived nuclides and application of k0-NAA to selected trace element, Nuclear Instruments and Methods in Physics Research A, 680, 2012.
[11]. F. De Corte, The k0-standardization method: a move to the optimization of neutron activation analysis, Habilitation Thesis, Univ. of Gent, 1987.
[12]. R. Acharya, J. Holzbecher, A. Chatt, Determination of k0-factors of short-lived nuclides and application of k0-NAA to selected trace elements, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 680, 1-5, 2012.
[13]. R. Acharya, A. Chatt, Characterization of the Dalhousie University SLOWPOKE-2 reactor for k0-NAA and application to medium-lived nuclides, Journal of Radioanalytical and Nuclear Chemistry, 257, 525-529, 2003.
[14]. H.M. Dung, P.D. Hien, The application and development of k0-standardization method of neutron activation analysis at Dalat research reactor, Journal of Radioanalytical and Nuclear Chemistry, 257, 2003.
[15]. H.M. Dung, Quality evaluation of the k0-standardized neutron activation analysis at the Dalat research reactor, Journal of Radioanalytical and Nuclear Chemistry, 309, 135-143, 2016.
[16]. R. Acharya, K. Swain, A. Kumar, N. Ajith, R. Verma, A. Reddy, Determination of k 0-factors and validation of k0-INAA for short-lived nuclides, Journal of Radioanalytical and Nuclear Chemistry, 286, 507-511, 2010.
[17]. F.D. Corte, A. Simonits, Recommended nuclear data for use in the k0 standardization of neutron activation analysis, Atomic Data and Nuclear Data Tables, 85, 2003.
[18]. H.M. Dung, Determination of k0 and Q0 factors for 5 short-lived radionuclides of interest in k0-NAA, Journal of Radioanalytical and Nuclear Chemistry, 2022.
[19]. M. Kubešová, k0-standardization in neutron activation analysis at LVR-15 reactor in Řež, 2012.
[20]. M.D. Ho, V.D. Ho, Q.T. Tran, N.S. Pham, V.C. Trinh, Q.D. Tran, T.S. Nguyen, M. Blaauw, The k0-standardized method for cyclic, epithermal and prompt gamma neutron activation analysis, 2017.
[21]. k0-database 2020. (n.d.). Kayzero.com. Retrieved December 18, 2022, from http://www.kayzero.com/k0naa/k0naaorg/Nuclear_Data_SC/Entries/2020/8/24_Update_of_k0-database_I-128.html.